logo
U.K. China Taiwan Home Site Map
About Us News ˛Product Support Contact Us
Product
Impedance Analyzers
line
spacer
Component Analyzers
line
spacer
Magnetics Analyzers
line
spacer
DC Bias Test Units
line
spacer
HF LCR Meters
line
spacer
LCR Meters
line
spacer
Telecom Test Units
line
spacer
Capacitance Meters
line
spacer
Accessories
line
spacer
•  Products by NF Corp New
line
spacer

Pay Low Yen for
the Best Products
from Japan

spacer
spacer ad4
distributor links
spacer

Education
Wayne Kerr makes it easy to
bring high quality measurement
products into your university.

Important Notices
img
icon Home icon Product iconDC Bias Test Units img
Product
spacer
 
tableline
tableline
line
line Zoomin line
line
6565 High Frequency DC Bias Units news
line
Features
  • Enhances usability of 6500 Analyzer
  • A single 6565 DC Bias Unit can deliver between 1 mA and 1 A of DC bias current in steps of 1 mA, and 1 A to 10 A in steps of 10 mA.
  • Up to four 6565 units can be paralleled to provide bias currents up to 40 A
  • 6565 Bandwith options of 5 MHz, 30 MHz and 120 MHz
  • Two basic fixtures available, 1026 (20 A) and 1027(40 A)
  • 1031 SMD adaptor available for 1026 and 1027 fixtures
line
Request for Quote PDF Download:Prod3200BSeries.pdf Information Request
tableline
tableline
tableline
tableline
To evaluate components at currents up to 40 A, the 6565 DC Bias Units are used with the Wayne Kerr 6500 series of precision Analyzers. When one 6565 DC Bias Unit is connected to an instrument, up to 10A of DC bias current can be selected. For currents up to 1A this current can be set in steps of 1mA and for currents from 1A to 10A the setting steps are 10mA. These step values multiply up in proportion to the number of bias units connected. By choosing appropriate models of 6500 and 6565, the user can obtain DC bias currents up to 40A at the full 120MHz operating frequency of the Analyzer.
Technical Specifications:
Measurement Function The 6565 series of DC bias units must be used in conjunction with either the 6500B Precision Impedance Analyzer or the 6500P HF LCR Meter.
Frequency Range 6565-05 20 Hz to 5 MHz
6565-30 20 Hz to 30 MHz
6565-120 20 Hz to 120MHz
Compliance Voltage 9V maximum at 1 V ac drive level
10V maximum at 0.25 V ac drive level
Basic Accuracyy ±1%, Varies with measurement speed, frequency and options chosen.
Measurement Fixtures 1026 is applicable for currents up to 20 A and the 1027 for currents up to 40 A. For SMD use the 1031 adaptor must be used in addition to the selected 1026 or 1027
General Data:
Input Specification Universal 85 to 265 V AC, 47 to 63 Hz
Input current 4.2 A rms max.
Power factor Meets EN61000-3-2
Unit powers up automatically when connected to a powered
analyzer. Isolating switch provided
Measurement Connections I2C bus link controls application of DC current and monitors status of analyzer. Status data includes excessive voltage drop and over temperature